Single flame Atomic Absorption Spectrometer AAS

1.The detection method and theoretical detectable trace elements of atomic absorption spectrometer: (for reference only)
1. Atomic absorption flame method
Atomic absorption flame method (air acetylene) can detect elements up to PPM level. Sodium (Na), potassium (K), magnesium (Mg), calcium (Ca), chromium (Cr), manganese (Mn), iron (Fe), cobalt (Co), nickel (Ni), platinum (Pt), gold (Au), copper (Cu), silver (Ag), zinc (Zn), cadmium (Cd), lead (Pb), bismuth (Bi), etc.
2. Atomic absorption hydride method
Atomic absorption hydride generation method can be used for trace detection of low melting point elements (PPb level). Measurement: Arsenic (As), Selenium (Se), Antimony (Sb), Bismuth (Bi), Lead (Pb), Tin (Sn), Tellurium (Te), Germanium (Ge), Mercury (Hg), etc. (Need to purchase a hydride generator separately)
2.Application scope of atomic absorption spectrometer
Atomic absorption spectroscopy is widely used for analysis and testing in various industries such as hospitals, pharmaceuticals, steel, health and epidemic prevention, metal smelting industry, geology and mineral resources, chemical industry, water quality monitoring, food and dairy products, environmental monitoring, quality inspection, drug testing, agriculture, toys, electronics, etc.
Specifications
| Model |
AAS-2001F8/F6/F3 |
| 1. Spectral system |
| Grating engraved lines |
1800 lines/mm |
| wavelength range |
190-900nm |
| spectral bandwidth |
Eight levels of automatic adjustability: 0.1, 0.2, 0.4, 0.6, 0.7, 1.0, 1.4, and 2.0 nm |
| wavelength accuracy |
≤0.15nm, Wavelength repeatability: ± 0.05nm |
| Baseline drift |
Static ≤ 0.002A/30 minutes, dynamic ≤ 0.004A/30 minutes |
| light source |
Automatic switching turret with eight light positions, automatic alignment |
| High performance power supply |
Built in high-performance lamp dual cathode power supply is used to measure edge energy spectral line interference and unstable elements such as arsenic, selenium, mercury, etc. High performance dual cathode luminescence improves luminescence intensity, prolongs element lamp service life, and enhances instrument testing stability and sensitivity |
| Instrument light source light bracket turret |
Adopting dual synchronous wheel technology control to improve accuracy and ensure smooth and reliable operation of the lighthouse |
| External optical path of the instrument |
By using a focusing type semi transparent and semi reflective device, the effective light source of the background is accurately combined with the main light path to improve the accuracy and stability of data analysis |
| Monochromator |
By using a worm gear transmission method to control the spectral system, the resolution accuracy of the optical path is greatly improved, and the wavelength accuracy of scanning the measured elements is also greatly improved |
| 3.Flame atomizer |
| Characteristic concentration (Cu) |
0.015μg/mL/1% |
| Detection limit (Cu) |
0.003μg/mL |
| precision |
RSD≤0.5% |